HY8200 User’s Manual
43
CURR:SLEW 1 // Set the rising rate for the current to 1 A/us
[SOURce]:CURRent:SLEW:RISE? // Query the rising rate of the current
7.6.12 [SOURce]:CURRent:SLEW:FALL
[SOURce]:CURRent:SLEW:FALL Set the falling rate for the current
[SOURce]:CURRent:SLEW:FALL <NRF+>
CURR:SLEW 1 // Set the falling rate for the current to 1 A/us
[SOURce]:CURRent:SLEW:FALL? // Query the falling rate of the current
7.6.13 "[SOURce]:VOLTage:SLEW
[SOURce]:VOLTage:SLEW Set the rising and falling rate for the voltage
[SOURce]:VOLTage:SLEW[:BOTH] <NRF+>
VOLT:SLEW 0.3 // Set the rising and falling rate for the voltage to 0.3 V/ms
[SOURce]:VOLTage:SLEW?
// Query the rising and falling rate of the voltage
7.6.14 [SOURce]:CURRent:PROTection
[SOURce]:CURRent:PROTection Set the protective value for the current
[SOURce]:CURRent:PROTection[:LEVel] <NRF+>
CURR:PROT 3 // Set the protective value for the current to 3A
[SOURce]:CURRent:PROTection[:LEVel]?
// Query the protective value of the current
7.6.15 [SOURce]:POWer:PROTection
[SOURce]:POWer:PROTection Set the protective value for the power
[SOURce]:POWer:PROTection[:LEVel] <NRF+>