HY8200 User’s Manual
52
1~16,0~40,200~99999 ,| MINimum | MAXimum OFF|OF
CURRent | VOLTage | RESistance | POWer | OPEN | SHORT
LIST:PARAMeter:ITEM 0,CURR,2.0,1000,OFF,1.0, 2.0
// Set the serial number to 1, mode to CC, constant value to 2.0, timing
to1000ms, check is off, the lower limit is 1.0, the upper limit is 2.0 for the
list mode
[SOURce]:LIST:PARAMeter:ITEM?
// Query each parameter of list in list mode
7.6.47 [SOURce]: LIST:TEST:RESUlts?
[SOURce]:LIST:TEST:RESUlts? Return the test result of list mode
[SOURce]: LIST:TEST:RESUlts?
1~16 or no parameter (return all test result)
LIST:TEST:RESUlts? 1
or
LIST:TEST:RESUlts?
[SOURce]: LIST:TEST:RESUlts?
// Query each parameter of the list in the list mode
0,CURR,1.00,OFF,0.00,0.00,PASS;1,CURR,3.00,OFF,0.00,0.00,PASS;2,
CURR,4.00,OFF,0.00,0.00,PASS;3,VOLT,5.00,OFF,0.00,0.00,PASS;
7.6.48 [SOURce]: LIST:TEST[:STATe]?
[SOURce]:LIST:TEST[:STATe]? Return the test state of list mode
[SOURce]: LIST:TEST[:STATe]?
LIST:TEST? 1 or LIST:TEST?
[SOURce]: LIST:TEST? // Query the test state of list mode
7.7 CHANnel[:LOAD] Channel Control Subsystem(only for dual-channel)
CHANnel[:LOAD] Set the channel to receive the insturction
1~2 or CH1、CH2 、ALL (channel)
CHANnel[:LOAD]?
// Query the channel that receive the insturction
7.7.1 CHANnel:SHORtcut[:COMMand]