CANHEAD enhancements
CANHEAD A01185_11_X00_00 HBM: public 71
Maximum data rates at 250 kBaud and 500 kBaud:
No. of modules Max. rate at 250 kBaud Max. rate at 500 kBaud
1 300 S/s 300 S/s
2 150 S/s 300 S/s
3 100 S/s 200 S/s
4 75 S/s 150 S/s
5 60 S/s 100 S/s
6 50 S/s 100 S/s
7 37.5 S/s 75 S/s
8 37.5 S/s 75 S/s
9 30 S/s 60 S/s
10 30 S/s 60 S/s
11 25 S/s 50 S/s
12 25 S/s 50 S/s
The “500 kBaud" selection is permanently stored in ML74B.
There is a new menu item under “ML74"- “Set minimum number of CAN
HEADs". This information allows different CANHEAD lines to be treated
equally; measurement is then synchronous, even if the number of assigned
modules differs.