CANHEAD enhancements
CANHEAD A01185_11_X00_00 HBM: public 69
12 CANHEAD enhancements
12.1 Enhanced data rate (with MGCplus only)
The CANHEAD system works at 250 kBaud as standard. The maximum bus
expansion is 250 m. 3000 measured values/s can be transmitted. For ex
ample, the maximum data rate with 4 modules = 40 channels is 75 S/s
(3000 S/s / 40).
If the requirements below are met, there is the option to increase the baud rate
to 500 kBaud.
Requirements for increasing the baud rate to 500 kBaud:
S The firmware of the ML74B being used must be P5.18.
S The MGCplus must be equipped with a CP42 (firmware P4.50) or CP52 .
S All bus masters (ML71BS6, ML78B) must also be able to understand sig
nals that are twice as fast (ML78B: firmware 5.28).
S All the CANHEADs in the line must have hardware revision 1.40.
S The firmware version in the CANHEADs must be 2.16.
S Cables must be no longer than 100 m.
S A terminator at the last module must be used for cables of 10 m or longer.
S The software on the PC (catman or Assistant) must be up‐to‐date.
If all these conditions are met, a data rate of 500 kBaud can be set.
From version 3.5 release 5 of the MGCplus Assistant, there will be a sub‐menu
CAN baud rate" under the enhanced settings (right‐click on ML74B).