102
The table below shows storable data.
Recording
method
Type of data
Data stored in one record:
record 1
Manual
recording
Standard mea-
surement data
(Data when
neither [TC] nor
[STEP] is on)
Data No., year/month/day/
hour/minute/second (at the
end of resistance measure-
ment), elapsed time,
test voltage (setting),
actual output voltage, resis-
tance (last) / (after 15 sec.) /
(after 30 sec.) / (after 1 min.),
PI, DAR, user-set interval for
PI x 2, and, resistance at
user-set interval x 2
Temperature cor-
rection data
(Data when [TC]
is on)
Data No., year/month/day/
hour/minute/second (at the
end of resistance measure-
ment), elapsed time,
temperature, humidity,
test voltage (setting),
actual output voltage, resis-
tance (last),
Reference temperature,
Resistance after correction
, and,
table No.
Step voltage test
data
(Data when
[STEP] is on)
Data No., year/month/day/
hour/minute/second (at the
end of test),
Step time,
temperature,
humidity
,
test voltage (setting)
,
actual output voltage
x 5,
Resistance
x
5
Logging
recording
- - - - - -
year/month/day/hour/minute/
second (at the start of logging
recording), Measuring interval,
temperature, humidity
,
test voltage (setting)
,
actual output voltage
x
360
times,
Resistance
x 360 times
In step voltage test resistance measurements,
only the last measurement at the end of each
step is recorded.
Voltage measurement data cannot be recorded.
Temperatures are not stored as logging records.