155
Err02
When power is turned on for
the first time after the backup
battery is replaced, Err02
appears.
Reinstall the
battery.
If Err02 appears even after
the battery has been rein-
stalled, the life of the backup
battery has expired, the bat-
tery is faulty, or some other
cause exists.
This
requires
repair.
(Err03 to
Err05 may
be tempo-
rarily dis-
played
during dis-
charging
after mea-
surement,
but this does
not indicate
a malfunc-
tion.)
Err03
Voltage measurement error
has occurred.
Err04
Current measurement error
has occurred.
Err05
Temperature measurement
error has occurred.
Err06
The discharge circuit is faulty.
E11
Details:
The actual temperature for temperature cor-
rection exceeds the convertible range.
Action:
Perform temperature correction within the
temperature ranges specified in the tables in
Appendix 4 "Temperature Correction Table"
(page 163).