3.1 Pre-Operation Inspection
1
2
3
4
5
6
7
8
9
10
11
付録
索引
55
7. Check to see if the insulation resistance
reading of the tester is 20 M.
8. Stop insulation resistance measurement.
See 3.2.2 "Ending Measurement" (page 64).
9. Short-circuit the tips of the clips of the
red and black test leads of the tester.
10. Press the key to see if the test volt-
age setting is [5.00 kV].
11. Hold down the key for more than
one second to start insulation resistance
measurement.
12. Check to see if the insulation resistance
reading of the tester is 0.00 M.
If a problem exists, discontinue use of
the tester.
Insulation resistance
Voltage