4.1 Using Timer
80
3. Press the key to confirm the entry.
If the key is pressed without pressing
the key, the tester returns to a
standby state with the time unchanged.
When the timer is successfully set, the
[TIMER] indicator lights.
4. Holding down the key for longer
than one second generates a test voltage,
and measurement begins.
At the bottom of the screen, remaining
time to completion of measurement is
displayed.
5. After the set time has elapsed, the tester
automatically stops measurement.
If the key is pressed, the tester
immediately stops measurement regard-
less of the remaining time.
Elapsed time at the completion of mea-
surement is displayed at the bottom of
the screen.
When the timer is set, auto power off is
disabled.