4.3 Temperature Correction (TC)
1
2
3
4
5
6
7
8
9
10
11
付録
索引
91
Press the keys shown in the table below to
switch the indication.
• If the resistance before correction
exceeds the measurement range, it
cannot be converted and the LCD
displays [
- - -].
• After the tester is placed in temperature
correction mode, measurement or input of
temperature and measurement of
insulation resistance may be conducted.
• However, if the tester is placed in
temperature correction mode when the
temperature is not held ( is
off), measure or enter temperature before
measuring resistance. You cannot
measure resistance first.
• Resistance measured by the step voltage
test cannot be converted using
temperature correction.
• In temperature correction mode, leakage
current may be displayed by pressing the
key but it cannot be corrected for.
Indications to be switched Keys used
key
Temperature
Reference temperature
Elapsed time
key
key
Insulation resistance
(after correction)
Leakage current
(no correction)
Setup screen of actual mea-
surement temperature
Standby state