Hioki PW3198 - Appendix 5 Detailed Explanation of IEC Flicker and V10 Flicker

Hioki PW3198
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix 5 Detailed Explanation of IEC Flicker and DV10 Flicker
A
18
The IEC flicker function is based on international standard IEC61000- 4-15, "Flickermeter - Functional
and design specifications".
Appendix 5 Detailed Explanation of IEC Flicker
and V10 Flicker
To measure the IEC Flicker or V10 Flicker
Flicker calculation and IEC flicker filter settings are configured on the SYSTEM-
DF1
[MAIN]-F1 [MEASURE] screen.
See:"5.1 Changing Measurement Conditions" (p.55)
IEC Flicker Meter
A/D
conver-
sion
Waveform
data
Block 1
Input voltage adapter
AGC
Block 2
Square multiplier
Square
multi-
plier
Luminous
sensitivity
filter
Block 3
Weighing filter
Voltage waveform
Square
multi-
plier
Block 4
Average
smoothing
1st order
Smooth
averaging
filter
Block 5
Online statistical analysis
Class
cate-
gory
Pst cal-
cula-
tion
Plt cal-
cula-
tion
Long interval
flicker
Plt
Instantaneous
flicker value
S(t)
Short interval
flicker
Pst