Chapter 11 Specifications
149
11
(1) Resistance range switching function
(2) Number of measurement digits selection function
(3) Measurement Speed
(4) Power Line Frequency Setting
(5) Zero Adjustment
Functions
Mode AUTO/ MANUAL (fixed to MANUAL when comparator function is ON)
Default setting AUTO
Number of measure-
ment digits selection
5digits/ 4digits
Default setting 5digits
Setting FAST/ MED/ SLOW
Default setting SLOW
Operation Selects the line voltage frequency
Setting AUTO (50 or 60 Hz, auto-detect)/ 50 Hz / 60 Hz
Default setting AUTO (auto-detect upon power on and resetting)
Operation Cancels the internal offset voltage and the surplus resistance.
Setting ON/ OFF (clear) : for each range
Range setting function NORMAL/ TIGHT
Adjustment range Selectable using the range setting function
NORMAL: Within -3% f.s. to 50% f.s. for each range (warning displayed
when in excess of 3% f.s. for each range) (f.s. = 30,000 dgt.)
TIGHT: Within -3% f.s. to 3% f.s. for each range (f.s. = 30,000 dgt.)
Default setting Zero adjustment: OFF, Range setting function: NORMAL