3.2 Setting the Measurement Speed
33
3
The measurement speed can be set to FAST, MED (medium), or SLOW.
The MED (medium) and SLOW settings offer increased measurement precision compared
to the FAST setting as well as greater resistance to the effects of the external environment.
If the setup is excessively susceptible to the effects of the external environment, shield the
measurement target and measurement leads adequately and twist the cables together.
See: "Appendix 7 Unstable Measured Values" (p.13)
Relationship Between Measurement Range and Speed
3.2 Setting the Measurement Speed
Press this to change the measurement speed.
Measurement speed
FAST
MEDIUM SLOW
50 Hz 60 Hz
Measurement time 21 ms 18 ms 101 ms 401 ms
With TC ON, comparator ON, and error of ±10%±2 ms
Integration time (detected voltage data acquisition time) reference values
FAST (50 Hz): 20.0 ms, FAST (60 Hz): 16.7 ms, MEDIUM: 100 ms, SLOW: 400 ms