Appendix 16 Calibration
A33
Appendix
Temperature measurement (Thermistor)
If the FLUKE 5520A cannot be used, please use the following equipment.
Calibration points
Equipment Calibration point Manufacturer Standard model
Multi-product calibrator 25°C, 2186.0 FLUKE Equivalent to 5520A
Equipment Calibration point Manufacturer Standard model
Dial-type resistor 25°C, 2186.0 Alpha Electronics Equivalent to ADR-6105M
Range Calibration points
Resistance measurement
30 m 0 , 10 m
300 m 0 , 100 m
3 0 , 1 or 3
30 0 , 10 or 30
300 0 , 100 or 300
3 k 0 , 1 k or 3 k
30 k 0 , 10 k or 30 k
300 k 0 , 100 k or 300 k
3 M 0 , 1 M or 3 M
Temperature (thermistor) 25°C: 2186.0 input