4.1 Zero Adjustment
46
2
Verify that the measured value is within ±3%f.s. If the zero-adjustment range
is set to NORMAL (-3%f.s. to 50%f.s.), zero-adjustment can be performed
when the measured value is 50%f.s. or less in each range, but a warning will
be issued when it is greater than 3%f.s.
If no measured value is displayed, verify whether the measurement leads
have been wired properly.
3
Perform zero-adjustment.
4
A confirmation message will be displayed. Confirm and return to the Mea-
surement screen.
Proper wiring Improper wiring
[0ADJ]
Perform zero-adjustment.
Switch the function menu
to P.2/2.
1
2
Perform zero-adjustment
and return to the Measure-
ment screen.
Cancel the operation and
return to the previous
screen.