4.4 Correcting Measured Values and Displaying Physical Properties Other than Resistance
55
4
2
Open the Measurement Settings Screen.
3
Enable the scaling function.
4
Set the gain coefficient.
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
Enables the scaling func-
tion
Disables the scaling func-
tion (default) (go to step 8)
Selection
2
1
Setting range: 0.2000×10
-3
to 1.9999×10
3
It is not possible to set the expo-
nent (E+3, etc.) directly. Use
and to multiply by 10
and 1/10 as necessary.
Move the cursor to the setting you
wish to configure. Make the value
editable with the key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
Multiply by 10.
Multiply by 1/10.
Clear value.