Evaluation - Assessment of roller bearing condition EasyBalancer EB 3500
10 - 8
Particularly when measuring the gSP value, care must be taken
to ensure that the measurement is always taken under the same
conditions (measurement setup, sensor mounting, measurement
location and direction). If possible, measurements should be tak-
en in the radial plane of the bearing, Fig. 10-20:.
Fig. 10-20: Favourable measuring point (A) and unfavourable measuring
points (B,C,D) for determining the gSP value.