TECHNICAL SPECIFICAT
IONS
EnCal 3000
Hardware
Manual
73022344 o
107
Analytical
Component
Range
Concentrations outside this scope might be possible, but verification
with Honeywell is needed
for natural gas
N2 : 0.005 – 20 %
CH4 : 55 – 100 %
CO2 : 0.001 – 20 %
C2 : 0.001 – 14 %
C3 : 0.001 – 6 %
C4 : 0.001 – 3 %
Neo-C5 : 0.005 – 0.25 %
C5 : 0.001 – 0.25 %
C6 : 0.001 – 0.1 %
C7 : 0.001 – 0.05 %
C8 : 0.001 – 0.05%
C9+ : 0.001 – 0.05 %
for biogas
N2 : 0.005 – 15%
CH4 : 60 – 100%
CO2 : 0.001 – 10%
C2 : 0.001 – 14 %
C3 : 0.001 – 6 %
C4 : 0.005 – 3 %
H2 : 0.001 – 5 %
O2 : 0.005 – 4%
H2S : 0.0002 – 1%
COS : 0.0001 – 1%
Uncertainty
< 0.1 % for all calculated properties
Repeatability
< 0.005 % (RSD) for all calculated properties :
Detection limit for C5: 1 ppm; Detection limit for C9: 2 ppm
Ambient conditions
Ambient
pressure
Atmospheric pressure
Temperature
range
for devices with ATEX/IECEx-Approval: 0 °C to 60 °C (32 to 140 °F),
extendable to -40 °C to 60 °C (-40 to 140 °F)
for devices with FM-Approval: 0 °C to 55 °C (32 to 130 °F),
extendable to -40 °C to 55 °C (-40 to 130 °F)
Humidity 0 – 100% RH

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell ENCAL 3000 and is the answer not in the manual?

Honeywell ENCAL 3000 Specifications

General IconGeneral
BrandHoneywell
ModelENCAL 3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals