EnCal 3000 Hardware Manual 73022344 o 97
8.3 Consumables (carrier and calibration gas cylinder)
A gas cylinder with the calibration gas is required to ensure that the
enCal 3000
operates
correctly. A gas cylinder containing helium (and or argon) to act as the carrier gas is also
required for operation. These gas cylinders should be replaced as soon as the outlet
pressure can no longer be held at a stable level. Depending on the regulator, this should
be around 1.0 MPag. At a pressure below 2 MPa, you should keep a replacement cylinder
available. The required work should be carried out using the steps described in
“
Connecting and replacing gas cylinders
”.
Hints about the needed gas cylinders could be found in the technical
specifications
8.3.1 Spare and replacement parts for general maintenance work
The device requires hardly any consumables and has hardly any wear parts. Honeywell
therefore recommends the annual inspection primarily for preventive reasons and to
replace the auxiliary gases if they are close to being empty and to identify errors. Our
contact details can be found at the start of this manual. Spare and replacement gases
can also be ordered using these contact details.
ATTENTION
For safety reasons, work which requires the device to be opened may only
be carried out by the manufacturer or by personnel trained or authorized
by it. Please contact us if the device suffers any damage or to arrange an
appointment for service work.
ATTENTION
Do not take any safety risks by carrying out incorrect maintenance work
or using non-genuine spare and replacement parts. No liability can be
accepted for damage caused by the use of non-genuine, non-approved
spare and replacement parts or incorrectly completed maintenance work.