MAINTEN
ANCE AND REPAIR
EnCal 3000
Hardware
Manual
73022344 o
87
CAUTION
The housing may only be opened by authorized/trained
personnel
You require a written operator license.
Use only explosion-proof, approved equipment.
Do not open the device without having taken all the necessary
safety precautions.
Do not disconnect cable inlets in a potentially explosive
atmosphere.Do not loosen or unscrew the breather valve.
Only carry out work after disconnecting the power supply and
depressurizing the system, the gas atmosphere must not be
potentially explosive.
Protect all open pipelines and connections from dirt.
Before the device is switched on again, ensure that all the gas
connections are sealed and the earthing or PE wire is correctly
installed.
.