6.3. MC Toolkit application software troubleshooting scenarios
136 MC Toolkit (MCT404) User Manual Revision 3
Could not download the selected
configuration to the Device due to any of the
following reasons:
Device mismatch
Version incompatibility
Communication errors
Try downloading to the
right device
Version incompatibility found! Some
of the parameters in version X of the
device are not supported in version
Y of the device. Download cannot
proceed!
Download of configuration from the file with
firmware version X could not be downloaded
to the device with version Y of the device
where X > Y. This is because there are some
parameters in version X that are not
supported in version Y
Try downloading to the
right device
Device mismatch found! Please
make sure that you have connected
to the right device...
The device type you trying to download the
selected configuration does not match with
the device type of the configuration in the file
Try downloading to the
right device
Download completed successfully!
Do you want to proceed to the MC
TOOLKIT screen?
In Offline mode, after completion of the
Download, you get this option to go back to
MC TOOLKIT screen to do an Upload of the
newly downloaded configuration
Could not read Tag IDs from the file
On parsing the XML file, there was some
error and could not read the Tag IDs from the
file
Make sure the
configuration file is not
corrupted. Reinstall the
MC Toolkit if the
problem persists
Could not set Tag ID! Do you want
to proceed with the rest of the
download?
When the Write Tag ID command fails, you
get this message. You can abort the rest of
the download or continue
Could not set Line Filter! Do you
want to proceed with the rest of the
download?
When the Write Line Filter command fails,
you get this message. you can abort the rest
of the download or continue
Could not set one or all of the
following parameters: Sensor type,
Cold junction compensation, Break
Detection, Linearity! Do you want to
proceed with the rest of the
download?
When the Write of any of the parameters in
this list fails, you get this message. You can
abort the rest of the download or continue
Could not download LRV / URV. The
device will keep the values valid for
its Sensor limits
When the Write of LRV / URV fails due to
invalid range, you get this message indicating
the device will set the values valid for the
current sensor
Could not set Damping! Do you want
to proceed with the rest of the
download?
When the Write Damping command fails, you
get this message. You can abort the rest of
the download or continue
Could not set the Critical Status
Latching configuration! Do you want
to proceed with the rest of the
download?
When the Critical Status Latching command
fails, you get this message. You can abort the
rest of the download or continue