5
*RCL
Syntax
*RCL
<
wsp
><
location
>
0
location
5
Denition
An instrument setting from the internal RAM is made the actual
instrument setting.
You can recall user settings from locations 1-5. See \*SAV".
Location 0 contains the default setting, which is the same as
that obtained by
*RST
.
Example
OUTPUT 724;"*RCL 3"
*RST
Syntax
*RST
Denition
The reset setting (default setting) stored in ROM is made the
actual setting.
Instrument state: the instrument is placed in the idle state
awaiting a command.
The following are not changed:
HP-IB (interface) state
Instrument interface address
Output queue
Service request enable register (SRE)
Standard event status enable register (ESE)
The commands and parameters of the reset state are listed in
the following table.
Remote Commands 5-11