206
Chapter 7, Configure Screen
Field Descriptions
(Gen)-(Anl)
This field is used with the RF Offset field to specify the amount of frequency
offset between the RF Generator and RF Analyzer.
This field is not displayed when the
RF Display field is set to Chan (the offset is
automatically set when using channel tuning).
See Also
"Setting an RF Generator/Analyzer Offset" on page 79
(Gen)-(Anl)
(User Defined)
This field defines the receiver-transmitter frequency offset when using User
Defined channel operation.
Use a positive value when the radio’s receive frequency is higher than the transmit
frequency (such as 45 MHz)
Use a negative value when the radios receive frequency is lower than the transmit
frequency (such as
45 MHz).
Operating Considerations
The value of this field is only used if the RF Display field is set to Chan, and the
RF Chan Std field is set to USER-DEF.
See Also
Chan Space (User Defined)
field description, on page 204
RF Chan Std
field description, on page 209
RF Display
field description, on page 210