528
Chapter 24, Connector, Key, and Knob Descriptions
Key Descriptions
INSTRUMENT STATE Keys
[LOCAL] returns the instrument to manual control after HP-IB control is used.
• [RECALL] lists any instrument setups that were saved.
• [MEAS RESET] clears the measurement “history” for all of the instrument’s measure-
ment algorithms (such as the averaging function) to restart all measurements that are in
progress.
• [PRESET] restores most instrument settings to their factory default states, (although
most CONFIGURE screen changes are not affected). Instrument self-diagnostics are
not run at this time.
• ADRS ([SHIFT], [LOCAL]) displays the HP-IB address of the Test Set.
• SAVE ([SHIFT], [RECALL]) stores instrument setups.
See Also
"Saving and Recalling Instrument Setups" on page 71
.
SCREEN CONTROL Keys
[RX], [TX], [DUPLEX], [TESTS], MSSG, HELP, and CONFIG access several instru-
ment control and information screens.
• [PREV] accesses the previous screen.
• HOLD ([SHIFT], [PREV]) stops all measurements. Selecting HOLD again resumes
measurements.
• PRINT ([SHIFT], [TESTS]) prints the entire contents of the displayed screen, the time
and date, and any Print Title defined in the PRINT CONFIGURE screen.
See Also
"Printing A Screen" on page 66.