EasyManua.ls Logo

HP 8921A - Ch Offset; Ext TX Key; Input Atten

HP 8921A
601 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90
Chapter 4, Adjacent Channel Power Screen
Field Descriptions
Ch Offset
Use the channel offset field to enter the frequency difference between the Tune
Freq
setting and the center of the adjacent channels you want to measure. This is
an absolute value; only positive values can be entered. See figure 11 .
Ext TX key
This field controls a switch at the MIC/ACC connector. Use it to “key” an
external transmitter.
See Also
"MIC/ACC" on page 514
Input Atten
Input Attenuation sets the amount of input attenuation for the RF IN/OUT and
ANT IN connectors. This function controls two settings:
The upper field determines if you want the instrument to set the attenuation automati-
cally (Auto), or if you want to set the value manually (Hold).
The lower field displays the present attenuation value, and is used to set the desired at-
tenuation level when the upper area is set to Hold.
Operating Considerations
Input Attenuator autoranging can interfere with oscilloscope or signaling decoder
operation under certain conditions.
See "Input Atten," in chapter 17, on page 421
for additional information.