EasyManua.ls Logo

HP 8921A - Page 477

HP 8921A
601 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
477
Chapter 22, Tests Screen
TESTS (Pass/Fail Limits)
Check
This field is used to select whether the test will verify only upper, only lower,
both, or none of the specified limits.
Operating Considerations
Selecting both upper and lower limits increases test time, but may be required for
some tests.
Lower Limit
This field is used to set the lower limit to be compared with the measured results.
If the measured result is below this limit, the test will fail.
Main Menu
Selecting this field returns you to the TESTS (Main Menu) screen.
Print All
This field allows you to print the Test Set’s screen image.
Spec#
The Spec# field lists the order of selected pass/fail limits for the current
procedure.
Units
The Units column indicates the unit-of-measure used for the limits (%, dBm,
kHz, and so forth).
Upper Limit
This field is used to set the upper limit to be compared with the measured results.
If the measured result is above this limit, the test will fail.

Table of Contents

Other manuals for HP 8921A

Related product manuals