3-19
Load Clinical Software and Configure System
032-0330-EN Rev K
9. If the DEXIS i-CAT FLX Server Plugin was selected to be installed:
a. On Welcome screen, click Next.
b. On DEXIS configuration screen, use the default folder path. Click Next.
c. On Confirm Installation screen, click Next. Installation will start and a progress bar is
displayed.
NOTE: If a pop-up displays indicating that a service is in use, click Continue, then reboot
the computer when the software is loaded.
d. When Complete window is displayed, click Close.
10. On SmartScan STUDIO Installer Executive, check that all selected programs have been
installed, then click Close. Remove media.
NOTE: The ImageRoot folder must be accessible to all client computers. The folder should be
shared and any password protection removed to ensure clients can access it, following the steps
below. Contact Technical Support if site network setup does not allow this.
11. To share the Image Root Directory with all users:
a. Open Control Panel and select Folder Options.
b. On the General tab, enable Show all folders.
c. On the View tab, enable Show hidden files, folders, and drives. Click OK.
d. Navigate to ProgramData\Dental Imaging Technologies Corporation (or to the location
where the Image Root directory resides).
e. Right-click ImageRoot folder, and select Share with->Specific people from the pop-up
menu.
f. Enter or select Everyone, and click Add.
g. Ensure Permission Level is set to Read/Write for Everyone. If not, select Read/Write from
drop-down menu.
h. Click Share.
i. Click Yes if prompted to turn on network discovery and file sharing, then click Done.
j. Right-click on ImageRoot folder and select Properties.
k. On the Sharing tab, click Network and Sharing Center link.
l. Click arrow next to All Networks to display options.
m. Under Password protected sharing, select Turn off password protected sharing.
n. Click Save changes.
o. Close Properties window.
12. Check settings on SmartScan STUDIO Integration Services Configuration:
NOTE: The configuration settings entered on this window are broadcast to all client workstations
that are configured to this host computer. Any changes to the configuration here will change the
configuration of the client workstations pointing to the host computer.