MEDIAJET – Operating instructions V09 Operation
32 INTEGRA Biosciences AG
4.4.1 Adapt dish height parameter
Measure the height of a stack of 10 using a ruler.
4.4.2 Filling dishes continuously or discontinuously
The carousel has a capacity of 360 and 540 Petri dishes respectively. However,
the program parameter ‘Number of dishes’ can be set up to 9999 dishes to be
filled. If the number of Petri dishes programmed exceeding the capacity of the
carousel, the filled dishes must be removed and empty ones must be loaded
during the filling process. For this there are two possibilities: Continuous or
discontinuous filling.
Biplate
alignment
(option)
To fill Petri dishes with two
compartments (with automatic
alignment of the dividing wall,
has to be enabled in program
configuration, see “4.2
Adapt
system parameter” on
page 26).
ON
OFF
OFF
Turbo mode
(option)
To accelerate filling of Petri
dishes (has to be enabled in
program configuration, see
“4.2
Adapt system parameter”
on page 26).
ON
OFF
OFF
Program
parameter
Description Value
range
Basic
setting
ASSISTANCE
It is important to always adapt the program setting dish height to ensure perfect
functioning of MEDIAJET.
NOTICE
In case you are working with Petri dishes higher than 21 mm you need to
underlay the filling nozzle holder with the supplied Base plate.
Press PROGRAM SETTINGS, choose the
parameter dish height and press CHANGE.
Enter the measured value.
Press SAVE in order to save this value.
The program adjustment is now complete.