Auto Advance Feature
The Auto Advance Feature is available only for the A+B format. If the Auto Advance feature
is enabled, the selected pattern is delivered when you press the footswitch, after which the
system automatically advances to the next pattern.
The orientation is selectable in clockwise and counter-clockwise on the Physician
Preferences.
For example, if octants 1 and 2 are selected, the system delivers octants 1 and 2 when you
press the footswitch, and then it automatically advances to octants 3 and 4. When you press
the footswitch again, the system delivers octants 3 and 4, and then automatically advances to
octants 5 and 6, and so on.
System Automatically Advances Pattern Clockwise
when Auto Advance Feature is Enabled