Instructions for use ELECTROmatic TM and TMM/TMC
10 Troubleshooting
82 / 90
Malfunction Cause Remedy
Device malfunctions
(no display, the LED on
the rear of the device
flashes).
No voltage supply on
the control panel.
No connection to the
control panel.
Check/restore correct connection.
Replace the connection cord to the control
panel.
Replace the control panel.
If error persists, notify the service engineer.
LUX light is red. Incorrect LED poles.
Rotate LED or replace it, if applicable.
SAFEdrive icon flashes
yellow.
SAFEdrive alert.
Motor is overloaded.
Also refer to: Chapter
on Protection Function
SAFEdrive
If a handpiece from a third-party manufacturer
is used, deactivate SAFEdrive or use a high-
speed handpiece from KaVo.
If you use a high-speed handpiece from KaVo,
relieve the handpiece of its load for at least 2
seconds, i.e. lift it off the tooth.
As soon as the motor reaches maximum speed
again and the SAFEdrive icon turns yellow,
continue working as usual making sure to use
intermittent mode as the operating mode.
See also:
2 3.6 Technical specifications of the ELECTRO-
matic, Page 27
2 6.4.3 Protection function SAFEdrive, Page 59
SAFEdrive icon flashes
red.
SAFEdrive error.
The motor is stopped
automatically.
Motor is overloaded.
Also refer to: Chapter
on Protection Function
SAFEdrive
If a handpiece from a third-party manufacturer
is used, deactivate SAFEdrive or use a high-
speed handpiece from KaVo.
If you use a high-speed handpiece from KaVo,
remove the handpiece from the mouth of the
patient and proceed as follows:
Carefully check the handpiece head for:
- Temperature
- Damage
- Ease of operation
The motor can be restarted if there is no dam-
age or overheating evident. To do so, press
the foot control again.
Service the handpiece.
If damage or overheating is evident, replace
the handpiece or have it repaired.
Event E2 Foot control is being
actuated during start-
up.
Do not press the foot control when you switch
the unit on.
Event E3 No motor attached.
Connect the motor.
Event E4 Motor blocked.
Relieve the motor, stop and start on foot con-
trol.
Event E5 Automatic light and
motor shut-off during
continuous operation of
the motor.
Comply with defined operating mode.