Instructions for use ELECTROmatic TM and TMM/TMC
10 Troubleshooting
Malfunction Cause Remedy
Event E8 Saved data or settings
have been reset to
starting value.
Confirm message and check or correct the pro-
gram settings.
If error persists, notify the service engineer.
Event E12 No connection to mo-
tor.
Check/restore correct connection.
Replace motor cord.
If error persists, notify the service engineer.
Event E14 Motor overload.
Relieve the motor, stop and start on foot con-
trol.
All other events Internal system error.
Turn the unit off and on.
If error persists, notify the service engineer.
83 / 90