SCPI Reference Tables 14-9
Command Description
Default
Parameter
Ref
SCPI
:TEMPerature Path to configure temperature:
:TRANsducer <name> Specify transducer; TCouple or INTernal. TCouple Sec 2
:TRANsducer? Query transducer.
:TCouple Thermocouple (TC): Sec 2
[:TYPE] <type> Set TC type; J, K, T, E, R, S, B or N. J
[:TYPE]? Query TC type.
:RJUNction Reference junction: Sec 2
:RSELect <name> Set reference type; SIMulated or INTernal. INTernal
:RSELect? Query reference type.
:SIMulated <n> Specify simulated temperature in °C; 0 to 60. 23
:SIMulated? Query simulated temperature.
:RTEMperature? Query internal temperature. Sec 2
:NPLCycles <n> Set integration rate in line cycles (PLC); 0.01 to 60
(60Hz) or 0.01 to 50 (50Hz).
5 Sec 3
:NPLCycles? Query NPLC.
:APERture <n> Set integration rate in seconds; 166.67e–6 to 1
(60Hz) or 200e–6 to 1 (50Hz).
83.33
:APERture? Query Aperture.
:DIGits <n> Set display resolution; 4 to 7. 6 Sec 3
:DIGits? Query display resolution.
[:CHANnel1] Channel 1 temperature commands:
:REFerence <n> Specify reference (Rel) value for Channel 1; –328
to 3310
0 Sec 4
:STATe <b> Enable or disable Rel. OFF
:STATe? Query state of Rel.
:ACQuire Use input signal as Rel.
:REFerence? Query Rel value.
:LPASs Control analog filter for TEMP1: Sec 3
[:STATe] <b> Enable or disable analog filter. OFF
[:STATe]? Query state of analog filter.
:DFILter Configure and control digital filter: Sec 3
:WINDow <n> Specify filter window (in %); 0 to 10. 0.01
:WINDow? Query filter window.
:COUNt <n> Specify filter count; 1 to 100. 10
:COUNt? Query filter count.
:TCONtrol <name> Select filter type; MOVing or REPeat. MOVing
:TCONtrol? Query filter type.
[:STATe] <b> Enable or disable digital filter. ON
[:STATe]? Query state of digital filter
Table 14-7
SENSe command summary (cont.)