16-4 Return to Section Topics 6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Command Description
Default
parameter SCPI
:CALCulate5
:FORMat <name>
:STATe <b>
:STATe?
:DATA?
:CALCulate6
:FORMat <name>
:STATe <b>
:STATe?
:DATA?
Subsystem to control CALC5 (RATIO):
Select ratio math; C3C4 (CALC3/CALC4);
C4C3 (CALC4/CALC3).
Enable or disable CALC5 math.
Query state of ratio.
Request CALC5 data.
Subsystem to control CALC6 (DELTA):
Select delta math; C3C4 (CALC3 - CALC4);
C4C3 (CALC4 - CALC3).
Enable or disable CALC5 math.
Query state of delta.
Request to CALC5 data.
C3C4
OFF
C3C4
OFF
9
9
9
9
9
9
:CALCulate7
:DATA?
:LATest?
:FEED <name>
:FEED?
:LIMit[1]
:COMPliance
:FAIL <name>
:FAIL?
:STATe <b>
:STATe?
:FAIL?
:LIMit2
:COMPliance
:FAIL <name>
:FAIL?
:STATe <b>
:STATe?
:FAIL?
Path to CALC7 (limit tests):
Return all CALC7 results.
Return most recent CALC7 result.
Specify Limit 3 to 6 input path (CALCulate3,
CALCulate4, CALCulate5, or CALCulate6).
Query input path.
Path to channel 1 hardware (compliance limits).
Path to mode.
Set fail mode (IN or OUT) of compliance
Query fail mode
Enable or disable Limit 1 test.
Query state of Limit 1 test.
Returns result of Limit 1 test: 0 (pass) or 1 (fail).
Path to channel 2 hardware (compliance) limit.
Path to mode.
Set fail mode (IN or OUT) of compliance
Query fail mode
Enable or disable Limit 2 test.
Query state of Limit 2 test.
Returns result of Limit 2 test: 0 (pass) or 1 (fail).
IN
15 or 7
IN
15 or 7
OFF
9
9
9
9
9
9
:LIMitX
:UPPer
[:DATA] <n>
[:DATA]?
:LOWer
[:DATA] <n>
[:DATA]?
:PASS
:STATe <b>
:STATe?
:FAIL?
:CLIMits
Path to software limits. LIMitX = LIMit3...LIMit6.
3
Configure upper limit:
Specify upper limit (-9.999999e20
to 9.999999e20).
Query upper limit.
Configure lower limit:
Specify lower limit (-9.999999e20 to
9.999999e20).
Query lower limit.
Path to sorting mode pass:
Enable or disable Limit X test.
Query state of Limit X test.
Return result of Limit X test: 0 (pass) or 1 (fail).
Composite limits for Limit 1 through Limit 6
1
-1
OFF
9
9
9
9
9
9
9
9
9
9
Table 16-1 (cont.)