EasyManua.ls Logo

Keithley 6482 - Page 198

Keithley 6482
350 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6482-901-01 Rev. A / August 2012 Return to Section Topics 16-5
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual Section 16: SCPI Command Reference
Command Description
Default
parameter SCPI
:CLEar
[:IMMediate]
:AUTO <b>
:AUTO?
:CALCulate8
:FORMat <name>
:FORMat?
:DATA?
Clear test results:
Clear latest limit test result.
Enable or disable clearing of test results when
:INITiate command is sent.
Query state of automatic clear.
Subsystem to control CALC8 (channel 1 buffer stats):
Specify math format (MEAN, SDEViation,
MAXimum, MINimum, or PKPK).
Query math format.
Read math result of CALC8.
ON
MEAN
9
9
9
9
3
LIMitX = LIMit3, LIMit4, LIMit5, or LIMit6.
Table 16-2
DISPlay command summary
Command Description
Default
parameter SCPI
:DISPlay
:ENABle <b>
:ENABle?
:MODE <name>
:MODE?
[:WINDow[1]]
:TEXT
:DATA <a>
:DATA?
:STATe <b>
:STATe?
:DATA?
:ATTRibutes?
:WINDow2
:TEXT
:DATA <a>
:DATA?
:STATe <b>
:STATe?
:DATA?
:ATTRibutes?
:DIGits<n>
:DIGits?
Turn on or turn off front panel display.
Query state of display.
Select display function. Name = CALC3,
CALC4, CALC5, CALC6, or DUAL.
Query selected display function.
Path to locate message to bottom display:
Control user test message:
Define ASCII message “a” (up to 32 characters).
Query text message.
Enable or disable message mode.
Query text message state.
Query data on bottom portion of display.
Query attributes of message characters: blinking (1)
or not blinking (0).
Path to locate message to bottom display:
Control user test message:
Define ASCII message “a” (up to 32 characters).
Query text message.
Enable or disable message mode.
Query text message state.
Query data on bottom portion of display.
Query attributes of message characters: blinking (1)
or not blinking (0).
Specify display resolution (4 to 7).
Query display resolution.
Note 1
DUAL
Note 2
Note 3
Note 2
Note 3
6
9
9
9
9
9
9
9
9
9
9
9
9
9
9
Notes:
1. *RST and :SYSTem:PRESet has no effect on the display circuitry. Pressing LOCAL or cycling power enables (ON) the
display circuit.
2. *RST and :SYSTem:PRESet has no effect on a user-defined message. Pressing LOCAL or cycling power cancels all
user-defined messages.
3. *RST and :SYSTem:PRESet has no effect on the state of the message mode. Pressing LOCAL or cycling power dis-
ables (OFF) the message mode.
Table 16-1 (cont.)
CALCulate command summary

Table of Contents

Related product manuals