17-36 Return to Section Topics 6482-901-01 Rev. A / August 2012
Section 17: Calibration Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Step 6. Channel 1 current calibration adjustment
Follow these steps to calibrate channel 1 current measurements. Table 17-14 summarizes
channel calibration currents and commands.
1. Connect the Model 6430 to the Model 6482 channel 1 INPUT connector, as shown in
Figure 17-9.
2. Select the Model 6430 V/R amperes function and set the output current to 1.90000 nA, and
then turn on the Model 6430 output.
3. Send the following command to select the channel 1, 2 nA range:
:SENS1:CURR:RANG 2e-9
4. Send the following command to calibrate the positive full-range current point:
:CAL:PROT:SENS1 1.9e-9
5. Set the Model 6430 current output to 0 nA.
6. Send the following command to calibrate the zero current point:
:CAL:PROT:SENS1 0
7. Set the Model 6430 current to -1.90000 nA.
Table 17-14
Calibration currents and commands
Connections Current range
Model 6430
calibration
currents
Model 5700A
calibration
currents Calibration commands
Channel 1:
Figure 17-9
Channel 2:
Figure 17-11
2 nA
+1.90000 nA
0.00000 nA
-1.90000 nA
-
-
-
:SENS1:CURR:RANG 2e-9
:CAL:PROT:SENS1 1.9e-9
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-9
20 nA
+19.0000 nA
0.0000 nA
-19.0000 nA
-
-
-
:SENS1:CURR:RANG 2e-8
:CAL:PROT:SENS1 1.9e-8
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-8
200 nA
+190.000 nA
0.000 nA
-190.000 nA
-
-
-
:SENS1:CURR:RANG 2e-7
:CAL:PROT:SENS1 1,9e-7
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-7
2 μA
+1.90000μA
0.00000μA
-1.90000μA
-
-
-
:SENS1:CURR:RANG 2e-6
:CAL:PROT:SENS1 1.9e-6
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-6
Channel 1:
Figure 17-10
Channel 2:
Figure 17-12
20 μA
-
-
-
+19.0000μA
0.0000μA
-19.0000μA
:SENS1:CURR:RANG 2e-5
:CAL:PROT:SENS1 1.9e-5
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-5
200 μA
-
-
-
+190.000 μA
0.000 μA
-190.000 μA
:SENS1:CURR:RANG 2e-4
:CAL:PROT:SENS1 1.9e-4
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-4
2 mA
-
-
-
+1.90000 mA
0.0000 mA
-1.90000 mA
:SENS1:CURR:RANG 2e-3
:CAL:PROT:SENS1 1.9e-3
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-3
20 mA
-
-
-
+19.0000 mA
0.0000 mA
-19.0000 mA
:SENS1:CURR:RANG 2e-2
:CAL:PROT:SENS1 1.9e-2
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-2