MMEMory:STATe:RECall:SELect <file>
MMEMory:STATe:RECall:SELect?
Selects which instrument state to use at power-on if the automatic recall mode is enabled
(MMEMory:STATe:RECall:AUTO ON). If the automatic recall mode is disabled (MMEMory:STATe:RECall:AUTO OFF), a
Factory Reset (*RST) is issued at power-on.
Parameter Typical return
<file> = Any valid file name "INT:\MyDAQ"
Select the power-down state to be used at power-on.
MMEM:STAT:REC:SEL "INT:\STATE_0"
MMEM:STAT:REC:AUTO ON
Save the current state into a file in the root directory of the internal flash file system and load it at power-on:
MMEM:STOR:STAT "INT:\MyDAQ"
MMEM:STAT:REC:SEL "INT:\MyDAQ"
MMEM:STAT:REC:AUTO ON
The specified folder must exist and cannot be marked as hidden or system.
The state file created by *SAV 0, called STATE_0.sta, resides in the root folder of the internal flash file system. It is
overwritten with the power-down state of the instrument when power is cycled.
When shipped from the factory, the instrument is configured to automatically recall the power-down state file
when power is restored.
This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
Keysight DAQ970A/DAQ973A Programming Guide 189
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish