[SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe] {OFF | ON} [, (@<ch_list>)]
[SENSe:]{RESistance|FRESistance}:POWer:LIMit[:STATe]? [(@<ch_list>)]
Enables or disables low-power for 2-wire and 4-wire resistance measurements.
Parameter is shared between RESistance and FRESistance measurements. Setting or querying the parameter
with the RESistance version of this command is identical to setting or querying it with the FRESistance ver-
sion.
This command is not supported in DMM Digitize data acquisition mode.
Error -221,"Settings conflict; DMM digitizing mode has been disabled" will be generated when you send the
command to enable low-power 2-wire or 4-wire resistance measurement in DMM Digitize mode.
Parameter Typical return
{OFF | ON}. Default: OFF 0 (OFF) or 1 (ON)
Configure 4-wire resistance measurements using a 10 PLC integration time. Enable low-power 4-wire res-
istance measurements. Then make and read one measurement:
CONF:FRES
FRES:NPLC 10
FRES:POW:LIM ON
READ?
Typical Response: +6.27530000E+01
– Low-power resistance measurements apply to the 100 Ω through 100 kΩ ranges only. The 1 MΩ through 1 GΩ
ranges source the same current regardless of the low-power setting.
– This parameter is set to its default value after a Factory Reset (*RST command).
244
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming