[SENSe:]{RESistance|FRESistance}:RANGe {<range> | MIN | MAX | DEF} [, (@<ch_list>)]
[SENSe:]{RESistance|FRESistance}:RANGe? [{(@<ch_list>) | MIN | MAX | DEF}]
Selects a fixed range for 2-wire and 4-wire resistance measurements.
Parameter is shared between RESistance and FRESistance measurements. Setting or querying the parameter
with the RESistance version of this command is identical to setting or querying it with the FRESistance ver-
sion.
Parameter Typical return
<range>: {100 Ω|1 kΩ|10 kΩ|100 kΩ|1 MΩ|10 MΩ|100 MΩ|1 GΩ}.
Default: 1 kΩ
+1.00000000E+03
Configure 2-wire resistance measurements, and select 10 kΩ range:
CONF:RES
RES:RANG
–
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
–
If the input signal is greater than can be measured on the specified fixed range, the instrument displays Overload
on front panel and returns "9.9E37" from the remote interface.
– For 4-wire resistance measurements, the instrument automatically pairs channel n in Bank 1 with channel n+10
in Bank 2 (DAQM900A or DAQM901A) or n+8 (DAQM902A) to provide the source and sense connections. For
example, make the source connections to the HI and LO terminals on channel 02 in Bank 1 and the sense con-
nections to the HI and LO terminals on channel 12 (DAQM900Aor DAQM901A) or 10 (DAQM902A) in Bank
2.Specify the paired channel in Bank 1 (source) as the <ch_list> channel (paired channels in Bank 2 are not
allowed in the <ch_list>).
– This parameter is set to its default value after a Factory Reset (*RST command).
Keysight DAQ970A/DAQ973A Programming Guide 245