[SENSe:]STRain:VOLTage:RANGe:AUTO {OFF | ON} [, (@<ch_list>)]
[SENSe:]STRain:VOLTage:RANGe:AUTO? [(@<ch_list>)]
This command is not supported in DMM Digitize data acquisition mode.
Error -221,"Settings conflict; DMM digitizing mode has been disabled" will be generated when you send the
command to enable autoranging for strain measurement in DMM Digitize mode.
Enables or disables autoranging for strain measurements. Autoranging is convenient because the instrument auto-
matically selects the range for each measurement based on the input signal detected.
Parameter Typical return
{OFF | ON}. Default: ON 0 (OFF) or 1 (ON)
– Autorange thresholds:
Down range at: <10% of range
Up range at:>120% of range.
– Selecting a fixed range (see [SENSe:]STRain:VOLTage:RANGe command) disables autoranging.
– This parameter is set to its default value after a Factory Reset (*RST command).