STATus:OPERation:CONDition?
Queries the condition register for the Standard Operation Register group. This is a read-only register and the bits
are not cleared when you read the register.The query reads the condition register and returns a decimal value that
corresponds to the binary-weighted sum of all bits set in the register.
Parameter Typical return
(none) +256
Reads the condition register (bit 8 is set): STAT:OPER:COND?
– The condition register bits reflect the current condition. If a condition is no longer true, the corresponding bit is
cleared in the condition register.
– A Factory Reset (*RST command) will set the "Configuration Changed" bit (bit 8) in the condition register.
STATus:OPERation:ENABle <enable>
STATus:OPERation:ENABle?
Enables bits in the enable register for the Standard Operation Register group. The selected bits are then reported to
the Status Byte. The query reads the enable register and returns a decimal value that corresponds to the binary-
weighted sum of all bits set in the register.
Parameter Typical return
Numeric. An integer from 0 to 65535. +512
Enables bit 9 in the enable register. STAT:OPER:ENAB 512
STATus:OPERation[:EVENt]?
Returns the sum of the bits in the event register for the Standard Operation Register group. This is a read-destruct-
ive register and these bits are cleared when you read the register. The query reads the enable register and returns a
decimal value that corresponds to the binary-weighted sum of all bits set in the register.
Parameter Typical return
(none) +512
Read the event register (bit 9 is set): STAT:OPER:EVEN?
– This register is cleared when it is read, or when you issue a *CLS command.