STATus:PRESet
Clears all enable register bits in Alarm Register, Standard Operation Register, and Questionable DataRegister.
Parameter Typical return
(none) (none)
Clear enable register bits: STAT:PRES
STATus:QUEStionable:CONDition?
Queries the condition register for the Questionable Data Register group. This is a read-destructive register and
these bits are cleared when you read the register. The query reads the enable register and returns a decimal value
that corresponds to the binary-weighted sum of all bits set in the register.
Parameter Typical return
(none) +2048
Reads the condition register (bit 11 is set): STAT:QUES:COND?
– The condition register bits reflect the current condition. If a condition is no longer true, the corresponding bit is
cleared in the condition register.
– A Factory Reset (*RST command) clears all bits in the condition register.
STATus:QUEStionable:ENABle <enable>
STATus:QUEStionable:ENABle?
Enables bits in the enable register for the Questionable Data Register group. The selected bits are then reported to
the Status Byte. The query reads the enable register and returns a decimal value that corresponds to the binary-
weighted sum of all bits set in the register.
Parameter Typical return
Numeric. An integer from 0 to 65535. +512
Enables bit 9 in the enable register. STAT:QUES:ENAB 512
STATus:QUEStionable[:EVENt]?
Returns the sum of the bits in the event register for the Questionable Data Register group. This is a read-destructive
register and the bits are cleared when you read the register. The query reads the enable register and returns a
decimal value that corresponds to the binary-weighted sum of all bits set in the register.
Parameter Typical return
(none) +514
Reads the questionable event register (bits 1 and 9 are set):STAT:QUES?
– Once a bit is set, it remains set until cleared by reading the event register or the *CLS (clear status) command.
– The *RST command has no effect on this register.
Keysight DAQ970A/DAQ973A Programming Guide 299
2 SCPI Programming