EasyManua.ls Logo

Keysight Technologies DAQ970A - Page 315

Keysight Technologies DAQ970A
393 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem:PERSona Subsystem
Command Summary
SYSTem:PERSona[:MANufacturer]
SYSTem:PERSona[:MANufacturer]:DEFault
SYSTem:PERSona:MODel
SYSTem:PERSona:MODel:DEFault
SYSTem:PERSona[:MANufacturer] {"<string>"}
SYSTem:PERSona[:MANufacturer]?
Sets the instrument's manufacturer ID string for backward compatibility.
Parameter Typical return
<string> =
{"Keysight Technologies" | ""Agilent Technologies" | "HEWLETT-
PACKARD"}
"Keysight Technologies", "Agilent Technologies", or "HEWLETT-
PACKARD"
Sets the instrument's manufacturer name to Keysight Technologies: SYST:PERS "Keysight Technologies"
The command will affect *IDN? that contains "Agilent Technologies", or "HEWLETT-PACKARD" instead of "Key-
sight Technologies".
SYSTem:PERSona[:MANufacturer]:DEFault
SYSTem:PERSona[:MANufacturer]:DEFault?
Returns the default manufacturer's ID string.
Parameter Typical return
(none) "Keysight Technologies"
Returns the default manufacturer's ID string: SYST:PERS:DEF
Keysight DAQ970A/DAQ973A Programming Guide 315