SYSTem:COMMunicate:LAN:UPDate
Stores any changes made to the LAN settings into non-volatile memory and restarts the LAN driver with the
updated settings.
Parameter Typical return
(none) (none)
Configures the instrument to use statically assigned LAN settings (disables DHCP):
SYST:COMM:LAN:DHCP OFF
SYST:COMM:LAN:DNS "198.105.232.4"
SYST:COMM:LAN:DNS2 "198.105.232.5"
SYST:COMM:LAN:GAT "198.105.232.1"
SYST:COMM:LAN:HOST "LAB1-DAQ970A"
SYST:COMM:LAN:IPAD "198.105.232.101"
SYST:COMM:LAN:WINS "198.105.232.4"
SYST:COMM:LAN:WINS "198.105.232.5"
SYST:COMM:LAN:UPD
Configures the instrument back to use DHCP (enables DHCP):
SYST:COMM:LAN:DHCP OFF
SYST:COMM:LAN.UPD
– Be very careful when you execute this command, because your instrument may not work on the LAN if you
update the instrument with invalid LAN settings.
– If your instrument does not work after you execute this command, perform the LAN Reset through instrument's
front panel softkey to restore the settings to reset values and reset the LAN, or use another I/O interface, such as
USB, to correct the settings.
– This command must be sent after changing the settings for DHCP, DNS, gateway, hostname, IP address, subnet
mask, or WINS.
334
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming