SYSTem:COMMunicate:LAN:WINS "<address>"
SYSTem:COMMunicate:LAN:WINS? [{CURRent | STATic}]
Assigns the static IP addresses of the Windows Internet Name System (WINS) servers. A primary and a secondary
server address () may be assigned. If DHCP is available and enabled, DHCP will auto-assign these server addresses.
These auto-assigned server addresses take precedence over the static addresses assigned with this command.
Contact your LAN administrator for details.
If you change this setting, you must execute a SYSTem:COMMunicate:LAN:UPDate command to activ-
ate the setting.
Parameter Typical return
<address> = Four-byte dot notation ("nnn.nnn.nnn.nnn"), where "nnn" in
each case is a byte value in the range 0 through 255.
[{CURRent | STATic}] Default: CURRent
"198.105.232.4"
Set a static primary WINS address:
SYST:COMM:LAN:WINS "198.105.232.4"
SYST:COMM:LAN:UPD
CURRent - Returns address currently being used by the instrument.
STATic - Returns address from non-volatile memory. This address is used if DHCP is disabled or unavailable.
SYSTem:SECurity:IMMediate sets this parameter to its default value.
The assigned DNS address is used for the DNS server if DHCP is disabled. Otherwise, the DNS server address is
auto-assigned by DHCP.
The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST) or after an