SYSTem:USB:HOST:ENABle {OFF | ON}
SYSTem:USB:HOST:ENABle?
Enables (On) or disables (Off) the front panel USB host port. Calibration security must be unsecured (CALibration:
SECure:STATe OFF,<code> command) before sending this command.
Parameter Typical return
{OFF|ON}
Default: OFF
0 (OFF) or 1 (ON)
Disables the front panel USB host port:
CAL:SEC:STAT OFF,MY_CAL_CODE
SYST:USB:HOST:ENAB OFF
CAL:SEC:STAT ON
If the front panel USB host port is disabled, it cannot be used to store or load measurements, state files, pref-
erences or other files or to perform firmware update.
The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST com-
mand), or after an Instrument Preset (SYSTem:PRESet command).