102 E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly–Level Problems
Self–Tests 27xx: A46 Offset Loop Self-Test Errors (Option UNY)
2700 Frac-N Signal Test
Replace the A46.
2701 Offset VCO Test
Replace the A46.
2702 Fstep OSL Test
Replace the A46.
2703 Offset Loop Detector Test
Replace the A46.
2704 Fcomb at YOL Test
Replace the A46.
2705 YO Loop Detector Test
Replace the A46.
Self–Tests 28xx: A8 Vector Output Self-Test Errors
2800 Power Supply Test
Replace the A8.
2801 Prelevel Loop Test
Replace the A8.
2802 ALC Mod System Test
Replace the A8.
2803 Pulse Mod System Test
Replace the A8.
2804 Clamp Detector Test
Replace the A8.
2805 Bias Switch Test
Replace the A8.