EasyManuals Logo

Keysight Technologies E8257D User Manual

Keysight Technologies E8257D
417 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
334 E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A26 MID (Microcircuit Interface Deck)
A26 MID (Microcircuit Interface Deck)
Tools Required
T–10 driver
Removal Procedure
Refer to Figure 3-46 and Figure 3-47 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument
Cover” on page 258 and Inner Instrument Cover” on page 260.
3. Position the signal generator with the A31 Motherboard facing you and the
RF deck on top.
4. If you have an Option 532, 540, or 544 signal generator, disconnect the
W29 ribbon cable from A26J33.
5. Disconnect the following ribbon cables:
W69 from A26J14
W28 from A26J15
W19 from A26J30
W36 from A26J13
W33 from A26J32
W34 from A26J31
W35 from A26J10
W20 from A26J20
W113 from A26J34 (E8257D & E8663D only)
W70 from A26J34 (E8267D only)
W99 from the A26J35 (Option 532 and 544 only)
W116 from the A26J36 (Option 532 and 544 only)
6. Disconnect the A25BW1 power cable from A26J12.
7. Using the T–10 driver, loosen the screw (1) that secures the A26 MID to
the RF deck.
8. Slide the A26 MID towards the front panel until the locking standoffs are
free.
9. Remove the A26 MID from the RF bracket.

Table of Contents

Other manuals for Keysight Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E8257D and is the answer not in the manual?

Keysight Technologies E8257D Specifications

General IconGeneral
BrandKeysight Technologies
ModelE8257D
CategoryPortable Generator
LanguageEnglish

Related product manuals