E8257D/67D, E8663D PSG Signal Generators Service Guide 289
Assembly Replacement
A13 I/Q MUX (E8267D Standard or with Option 015) (Serial Prefixes
<US4722/MY4722)
A13 I/Q MUX (E8267D Standard or with Option 015)
(Serial Prefixes <US4722/MY4722)
Tools Required
T–10 driver
needle–nose pliers
Removal Procedure
Refer to Figure 3-19 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument
Cover” on page 258 and Inner Instrument Cover” on page 260.
3. Position the signal generator with the front panel facing you.
4. Using the T–10 driver, remove the four screws from the Baseband
Generator cover.
5. Using the needle–nose pliers, disconnect the following flexible cables:
W63 from A13J5
W62from A13J6
W49 from A13J7
W50 from A13J8
W61 from A13J9
W60 from A13J10
W59 from A13J11
W58 from A13J12
W57 from A13J13
W56 from A13J14
6. Lift the retention levers and disconnect the A13 I/Q MUX from the A31
Motherboard.
7. Lift the A13 I/Q MUX out of the slot.
Replacement Procedure
1. Reverse the order of the removal procedure. (See the internal cover
overlay for cable color codes.)

Table of Contents

Other manuals for Keysight Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E8257D and is the answer not in the manual?

Keysight Technologies E8257D Specifications

General IconGeneral
BrandKeysight Technologies
ModelE8257D
CategoryPortable Generator
LanguageEnglish

Related product manuals