E8257D/67D, E8663D PSG Signal Generators Service Guide 121
Troubleshooting
Troubleshooting Performance Test Problems
External Pulse Modulation ON/OFF Ratio If the frequency is 3.2 GHz, check the
A11 Pulse/Analog Modulation Generator or the A8 Output.
If the frequency is >3.2 GHz, check the
A11 Pulse/Analog Modulation Generator and the A10 ALC.
Harmonic Spurious
Sub–Harmonic Spurious
Refer to “Troubleshooting Harmonic Spurious” on page 127.
Non–Harmonic Spurious Refer to “Troubleshooting Non–Harmonic Spurious” on page 132.
Single–Sideband Phase Noise Refer to “Troubleshooting the RF Path” on page 122.
Internal EVM Digital Modulation Quality Perform I/Q and digital calibrations.
Digital Modulation Power Relative to CW Perform I/Q and digital calibrations.
Performance test that is failing... Action to perform...