EasyManuals Logo

Keysight Technologies E8257D User Manual

Keysight Technologies E8257D
417 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
E8257D/67D, E8663D PSG Signal Generators Service Guide 87
Troubleshooting
Troubleshooting Assembly–Level Problems
3. The voltage measured by the DVM should be approximately 1.8 V. If it is
not, use an oscilloscope to view P131–6 of the A31 Motherboard (this is
the pulse signal out of the A11 Pulse/Analog Modulation Generator).
Change the pulse period to 1 ms and the pulse width to 500 μs.
If the pulse is not present, perform the A11 Pulse/Analog
Modulation Generator self–test, and follow the troubleshooting
procedure.
4. If pulse modulation is working, perform the A8 Output self–test.
If the self–test passes, refer to “Troubleshooting the RF Path” on
page 122, and check the lowband (<3.2 GHz) path.
If the self–test fails, refer to the A8 Output test troubleshooting.
1206 20 GHz IQ Modulator
1. If the power supplies +8V AF, +2.5V AF, +9V AFSW, +12V AFSW, +15V
VLIMF, or 15VSWBIAS fail, refer to the table page 83, and check the
supplies used by the A35 I/Q Modulator as they leave the A26 MID
connector.
If the signals are good at the A26 MID connector, replace the A35
I/Q Modulator.
2. If the VD724 Doubler, test index 6, fails, replace the A35 I/Q Modulator.
3. If self–test TC672RF indexes 8, 9, 11, or 12 fail:
a. Remove the cable from J2 of the A29 20 GHz Doubler.
b. While measuring the signal at J2 with a spectrum analyzer, run
self–test 1206. You should see a +20 dBm signal at 5 GHz.
If the signal is bad, troubleshoot the RF path.
If the signal is good and TC672RF indexes 8, 11, or 12 have
failed, replace the A35 I/Q Modulator.
If the signal is good and TC672RF index 9 has failed, you must
also check the I/Q signals from the A13 I/Q Multiplexer:
1. Remove the cable from J6 (I) of the A35 I/Q
Modulator, and connect it to an oscilloscope.
2. Execute a single 1206 self–test. The signal on the
oscilloscope should pulse to approximately 0.5 Vdc
for approximately 500 ms.
If the signal is bad, refer to the A13 I/Q
Multiplexer troubleshooting.
3. If the signal is good, reconnect the cable to J6,
remove the cable to J7 (Q), and connect this cable
to the oscilloscope.

Table of Contents

Other manuals for Keysight Technologies E8257D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E8257D and is the answer not in the manual?

Keysight Technologies E8257D Specifications

General IconGeneral
BrandKeysight Technologies
ModelE8257D
CategoryPortable Generator
LanguageEnglish

Related product manuals