Loading...
90 E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly–Level Problems
1300 Power Supply
Replace the A11 Reference.
1301 Internal Pulse Generator Clock
1. Using an oscilloscope, measure P131–91 for a 10 MHz signal at 3.5 V
pp
.
If the signal is good, replace the A11 Pulse/Analog Modulation
Generator.
2. If the signal is bad, check P41–7.
If the signal is good on P41–7, replace the A31 Motherboard.
If the signal is bad on P41–7, replace the A7 Reference.
1302 Output
Replace A11 Pulse/Analog Modulation Generator.
1303 Voltage Ref DAC
Replace A11 Pulse/Analog Modulation Generator.
1304 20 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1305 Not Used
1306 Standard 3 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1307 High Performance 3 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1308 Numeric Synthesizer
Replace A11 Pulse/Analog Modulation Generator.
+5.2 Digital high P131–64, 65, 129,
130
+5.04 +5.36 Main Supply
+3.4 Digital Low P131–60, 61, 62, 63,
125, 126, 127, 128
+3.29 +3.5 Main Supply
Supply Voltage
(Vdc)
Connector Pins Minimum Value
(Vdc)
Maximum Value
(Vdc)
Origin